U
?ID
?SN
?TN
?PT
PT: ∗∗∗∗∗. ∗

15. COMMANDS

15-1 Command List

Note

A command has a terminator added, that is specified using the "Terminator (Crlf)" parameter of “Serial interface ( 5if )” in the function table, and is sent to the balance.

Commands to query weight data

C

Cancels the S or SIR command.

 

 

Q

Requests the weight data immediately.

S

Requests the weight data when stabilized.

 

 

SI

Requests the weight data immediately.

SIR

Requests the weight data continuously.

ESCP

Requests the weight data when stabilized.

 

 

Note: The “Q” and “SI” commands, the “S” and “ESCP” commands behave the same.

Commands to control the balance

CAL

Same as the

CAL

key.

 

 

 

 

 

 

OFF

Turns the display off.

 

 

 

 

 

 

ON

Turns the display on.

P

 

 

 

 

 

 

 

Same as the

ON:OFF

 

key

 

 

 

 

 

PRT

 

 

 

 

 

 

 

Same as the

PRINT

key

 

 

 

 

R

 

 

 

 

 

 

 

Same as the

RE-ZERO

key

 

 

 

SMP

 

 

 

 

 

 

 

Same as the

SAMPLE

key.

 

 

T

Tare key

 

 

Z

 

 

 

 

 

 

 

Same as the

RE-ZERO

key

 

 

ESCT

 

 

 

 

 

 

 

Same as the

RE-ZERO

key

Same as the MODE key

Requests the ID number.

Requests the serial number.

Requests the model name.

Requests the tare weight.

g Changes the tare weight.

The unit added is the current weighing unit in A&D standard format. Note: The “R”, “Z” and “ESCT” commands behave the same.

ESC : ASCII code 1Bh

51

Page 53
Image 53
A&D FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i instruction manual Command List, Commands to query weight data