9-4 Description of the Class “Environment, Display”

Condition ( Cond )

Cond 0

This parameter is for sensitive response to the fluctuation of a weight value.

 

Used for powder target weighing, weighing a very light sample or when

 

quick response weighing is required.

 

 

 

 

 

 

 

 

After setting, the balance displays

FAST.

 

 

 

Cond 2

This parameter is for stable weighing with slow response. Used to prevent a

 

weight value from drifting due to vibration or drafts.

 

 

 

 

 

 

After setting, the balance displays

SLOW.

 

 

Note

 

With “Hold function (Hold)” set to “ON (1)”, this item is used to set the

 

averaging time.

Stability band width ( 5t-b)

This item controls the width to regard a weight value as a stable value. When the fluctuation per second is less than the parameter, the balance displays the stabilization indicator and outputs the data. The parameter influences the “Auto print mode”

5t-b 0 This parameter is for sensitive response of the stabilization indicator. Used for exact weighing.

5t-b 2 This parameter ignores slight fluctuation of a weight value. Used to prevent a weight value from drifting due to vibration or drafts.

Note

With “Hold function (Hold)” set to “ON (1)”, this item is used to set the stabilization range.

Hold function ( Hold ) (Animal weighing mode)

This function is used to weigh a moving object such as an animal.

When the weight data is over the weighing range from zero and the display fluctuation is within the stabilization range for a fixed period of averaging time, the processing indicator illuminates and the balance displays the average weight of the animal. When the animal is removed from the weighing pan, the display returns to zero automatically.

This function is available only when the hold function parameter is set to “1” (the animal mode indicator ANIMAL illuminates) and any weighing unit other than the counting mode is selected.

The stabilization range and averaging time are set in “Condition (Cond)” and “Stability band width (5t-b)”.

Weighing range

FX-120 i /200 i / 300 i

0.200 g or over

FX-1200 i /2000 i / 3000 i

2.00 g or over

 

 

Averaging time

Stabilization range

Cond 0

2 seconds

Faster

5t-b 0

Lesser

Cond

1

4 seconds

5t-b

1

Cond

2

8 seconds More accurate

5t-b

2

Greater

 

 

 

 

 

 

 

34

Page 36
Image 36
A&D FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i Description of the Class Environment, Display, Condition Cond