Class

Item and Parameter

 

 

Description

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

Key mode

 

Accepts the

PRINT

key only when

 

 

 

 

 

the display is stable.

 

 

 

 

 

 

 

 

 

 

 

1

Auto print mode A

 

Outputs data when the display is

 

 

 

 

 

(Reference = zero)

 

 

 

 

 

 

 

stable and conditions of ap-p, ap-b

 

 

 

 

 

Auto print mode B

 

 

 

 

 

2

 

and the reference value are met.

 

prt

(Reference = last stable value)

 

 

 

 

 

 

 

 

Data output mode

3

Stream mode

 

Outputs data continuously.

 

 

 

 

 

 

 

 

 

 

 

 

4

Key mode B (Immediately)

 

Accepts the

PRINT

key regardless

 

 

 

 

 

of the display condition.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Accepts the

PRINT

key immediately

 

 

 

 

5

Key mode C (When stable)

 

when the display is stable, or waits

 

 

 

 

 

 

 

for the display to be stable when not.

 

 

 

 

6

Interval output mode

 

Uses interval output mode.

 

 

 

 

0

Plus only

 

Displayed value>Reference

 

 

 

 

 

ap-p

 

1

Minus only

 

Displayed value<Reference

 

 

 

Auto print polarity

 

 

2

Both

 

Regardless of displayed value

 

 

 

 

 

 

ap-b

0

10 digits

 

Difference between reference value

 

1

100 digits

 

 

 

 

 

 

 

 

 

 

 

and displayed value

 

Auto print difference

 

dout

2

1000 digits

 

 

 

 

 

 

 

 

Data output

 

 

 

0

Every measurement

 

 

 

 

 

 

 

 

1

2 seconds

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

5 seconds

 

 

int

3

10 seconds

Interval time for the interval output

Interval time

4

30 seconds

mode

 

(With

prt 6

 

5

1 minute

)

 

 

 

 

6

2 minute

 

 

 

7

5 minute

 

 

 

8

10 minute

 

 

pU5e

 

0

No pause

Selects the data output interval.

 

 

Data output pause

1

Pause (1.6 seconds)

 

 

at-f

 

0

Not used

Selects whether or not automatic

 

 

Auto feed

1

Used

feed is performed.

info

 

0

No output

 

 

 

 

 

 

 

 

1

AD-8121 format

Selects GLP output method.

GLP output

2

General data format

 

 

 

 

 

ar-d

 

0

Not used

Adjusts zero automatically after data

 

 

Zero after output

1

Used

is output

Factory settings. Digit is a unit of minimum weighing value

32

Page 34
Image 34
A&D FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i Data output mode, Auto print polarity, Auto print difference