A&D FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i Outputting the statistical results

Models: FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i

1 72
Download 72 pages 26.17 Kb
Page 20
Image 20

1 Press the RE-ZERO key to set the display to zero.

2 Place the sample on the weighing pan and wait for the stabilization indicator to turn on.

3 Press the PRINT key to add the data displayed to statistical calculation. The number of data on the upper left of the display increases by 1.

4 Repeat steps 1 to 3 for each weighing.

Outputting the statistical results

5 Each time the MODE key is pressed, the display changes: the results as selected in “Statistical function mode output items (5taf)”, CanCel and Clear .

Notes

When the number of data is 1, the standard deviation and coefficient of variation are displayed as - - - - - - - .

When the average is 0, the coefficient of variation is displayed as - - - - - - - .

Statistical items are indicated on the upper left of the display using the following symbols.

Symbol

Statistical item

 

SUM

Sum

 

MAX

Maximum

 

MIN

Minimum

 

R

Range (Maximum – minimum)

 

AVE

Average

 

SD

Standard deviation

 

CV

Coefficient of variation

 

6 While the results are displayed, press the

 

PRINT key to output the results.

Output example

18

Page 20
Image 20
A&D FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i instruction manual Outputting the statistical results