Example

Example 1:

Example 2:

Example 3:

Example 4:

Example 1:

Example 2:

Example 3:

Example 4:

Notes:

Table C-3. Examples of Questionable Status Register Bit Programming

 

Signal

 

Cond Reg

 

PTR

 

NTR

 

Event

Enable

Summary

 

 

Change

 

Filter

 

Filter

 

Reg

Reg

Bit

OT

0

1

1

 

0

 

1

 

1

1

UNR

0=1or

1

 

1

 

1

 

1

 

RI

 

 

x

 

x

 

x

 

0

0

RI

 

 

0

 

1

 

0

 

1

1

Explanation of Examples

Allow an overtemperature event to set the QUES summary bit.

STAT:QUES:ENAB 16;PTR 16

Allow both an unregulated event and the removal of that event to set the QUES summary bit

STAT:QUES:ENAB 1024;PTR 1024;NTR 1024

Prevent any remote inhibit event from setting the QUES summary bit. STAT:QUES:ENAB 0 (or any value that does not set bit 9)

Allow only the removal of an existing remote inhibit event to set the QUES summary bit

STAT:QUES:ENAB 512;NTR 512

Boldface (1) indicates programmed state; italic (1) indicates signal state. "x’’ = ’’don’t care’ condition.

Register

Condition

PTR Filter

NTR Filter

Event

Enable

Table C-4. STAT:OPER Commands

Command

Query

(None)

STAT:OPER:COND?

STAT:OPER:PTR <NRf>

STAT:OPER:PTR?

STAT:OPER:NTR <NRf>

STAT:OPER:NTR?

(None)

STAT:OPER:EVEN?

STAT:OPER:ENAB <NRf>

STAT:OPER:ENAB?

Clear By

Cannot be cleared Programming 0 Programming 0 Reading or *CLS Programming 0

Status Programming Example. Control over Conditional Status events is the same as for Questionable Status events. See Table C-1 for register configurations.

STAT:OPER:ENAN 32;NTR 32

Allow the OPER summary bit to be set whenever the power supply leaves CV mode

Standard Event Status Group

Register Functions. The Standard Event Status group consists of two registers (Event and Enable) that are progammed by common commands, not by STAT subsystem commands.

The Standard Event register latches events relating to interface communication status (see Table C-1). It is a readonly register that is cleared when read.

Read Query

*ESR?

Clear Commands

*CLS *ESR?

The Standard Event Enable register is a read/write register that functions similarly to the enable registers of the Operation and Questionable status groups.

Program Command

*ESE <NRf>

Read Query

*ESE?

 

Clear Command

*ESE 0

 

Standard Commands for Programmable Instruments (SCPI) 107

Page 107