OUTP:PROT

There are two output protection commands that do the following:

OUTP:PROT:CLE Clears any OV (overvoltage), OC (overcurrent, unless set via external voltage control),OT (overtemperature), or RI (remote inhibit) status conditions. If the power supply is configured with the Agilent 59510A/llA Relay Accessory, it will be restored to the state it was in before the protection circuit tripped.

OUTP:PROT:DEL Sets the time that a change in CV, CC, or UNREG condition must appear at the output before that change is recorded in the Status Operation Condition register. The delay prevents momentary output changes that can occur during reprogramming from being registered as events by the status subsystem.

Command Syntax

OUTPut:PROTection:CLEar

 

OUTPut:PROTection:DELlay <NRf+>

Parameters

OUTP:PROT:CLE (None)

 

OUTP:PROT:DEL See "DI,Y" in Table 3-7 of Section III.

Default Sufflx

S

Examples

OUTP: PROT: CLE OUTPUT: PROTECTION: CLEAR

 

OUTPUT:PROTECTION:DELAY 75E-l

 

OUTP:PROT:DELMIN OUTPUT:PROT:DELAY MAX

Query Syntax

OUTP:PROT:CIE (None) OUTP:PROT:DEL?

 

OUTP:PROT:DEL? MIN OUTP:PROT:DEL? MAX

Returned Parameters

<NR3> OUTP:PROT:DEL returns value of programmed delay.

 

OUTP:PROT:DEL? MIN and OUTP:PROT:DEL? MAX return the minimum and maximum

 

programmable delays.

OUTP:REL[:STAT]

 

This command is valid only if the power supply is configured for the optional Agilent 59510A and Agilent 59511A Relay Accessory. Programming ON closes the relay; programming OFF opens it. If the power supply is not configured for either device, the command generates an error.

Command Syntax

OUTPut:RELay[:STATe] <bool>

Parameters

0 or OFF

I or ON

*RST Value

0

 

Examples

OUTP: REL 1 OUTPUT: RELAY OFF OUTPUT: RELAY: STATE OFF

Query Syntax

OUTP:REL?

Returned Parameters

<NR.1>

0 or 1

OUTP:REL:POL

This command is valid only if the power supply is configured for the Agilent 59511A Relay Accessory. The NORMal command causes the device output polarity to agree with that of the power supply. Programming REV causes the polarity of the device to be opposite to that of the power supply. If the power supply is not configured for the Agilent 59511A, either command generates an error.

Command Syntax

OUTPut:RELay:POLarity {NORMalREVerse}

Parameters

<CRD> NORMal or REVerse

*RST Value

NORM

Examples

OUTP: REL: POL NORM OUTPUT: RELAY: POLARITY REVERSE

Query Syntax

OUTP:REL:POL?

Returned Parameters

<CRD> NORM or REV

98Standard Commands for Programmable Instruments (SCPI)

Page 98
Image 98