Number

5

Table 3-1. Controls and Indicators (continued)

 

Controls/Indicators

Description

Page

 

 

 

Numeric Display

Two 3-1/2 digit alphanumeric displays with automatically

33

 

positioned decimal point that ordinarily indicate output VOLTS

 

 

and AMPS (see items 6 & 7). When power supply is turned on all

 

 

segments light for approximately 1 second. During an error

 

 

condition, power supply output may exceed display range; displays

 

 

will indicate + OL or - OL.

 

 

 

 

6

7

8

DISPLAYS SETTINGS Pushbutton Switch

DISPLAY OVP Pushbutton Switch

OUTPUT ADJUST Controls - Rotary Pulse Generator (RPG) and pushbutton switch

Causes numeric displays to indicate programmed voltage and

35

current values, rather than actual output values; allows both settings

 

to be made without the necessity of opening or shorting load.

 

 

 

Causes VOLTS display to indicate OVP trip voltage, AMPS

38

display is blanked; allows setting to be made without changing

 

output settings or load connections.

 

 

 

OUTPUT ADJUST knob functions either as a voltage control or a

38

current control, as determined by the pushbutton switch and

 

indicated by whichever (green) indicator, VOLTAGE or

 

CURRENT, is on. Knob functions as a two-speed device; faster

 

rotation causes greater rate of change per revolution. OUTPUT

 

ADJUST controls operate only when unit is under local control.

 

 

 

9

FOLDBACK Control

The pushbutton switch toggles foldback protection on and off in local operation; has no effect if power supply is not in CV or CC (ERROR LED flashes), or is in remote. Switch also resets foldback protection circuit if it has disabled power supply output. FOLDBACK ENABLED LED (green) operates in either local or remote.

39

10

11

OVP ADJUST

LINE Switch

The recessed, single-turn screwdriver control sets the overvoltage protection trip voltage.

Turns ac power on and off.

43

34

NOTE

Under certain conditions of line and load, it is possible for the supply to provide more than rated output

 

power and still maintain regulation. If this occurs, the unit will operate normally and the OVERRANGE

 

indicator will be off. However, the slightest change in either line or load may cause the unit to go out of

 

regulation. Operation of the unit beyond the rated-output-power boundary is not recommended under any

 

circumstances.

 

 

Turn-On Checkout Procedure

The power supply performs a series of self tests each time power is turned on. All front-panel LEDs, including all meter segments, are also turned on. The tests take approximately one second to complete, and all indicators remain on while the tests are running. This alerts the operator that self tests are running, and allows the operator to note if any indicators are inoperative.

After the self tests are completed all front-panel indicators are turned off for one-half second, allowing the operator to note if any are stuck on. If the operator suspects that any indicator may be malfunctioning he should turn power off and back on again while observing that indicator.

Operating Instructions 33

Page 33
Image 33
Agilent Technologies 6028A, 6035A, 6033A, 603xA, 6023A, 6038A 6010A, 6031A, 6032A manual Turn-On Checkout Procedure, OVP Adjust