Reset. Reset restores the power supply output if it has been disabled by OVP, foldback, or remote inhibit. The output returns to the present voltage and current settings; the values may be changed while the output is disabled. The power supply is reset with the command:
RST
Note that if the condition which caused OVP or foldback remains, the power supply output will be disabled again after reset. If a condition which tripped the INH input (set the RI bit true (1)) remains after a RST command is issued, the supply will not be
Hold and Trigger. The power supply contains
This feature allows synchronization of multiple instruments on the
Hold is turned on and off using these codes:
HOLD OFF or HOLD 0
HOLD ON or HOLD 1
The state of the hold on/off function may be read by sending:
HOLD?
and addressing the power supply to talk. The response from the power supply is in this format:
HOLD 0 or HOLD 1
in which 0 indicates that hold is turned off, and 1 indicates that hold is turned on.
Values that are held in first rank can be shifted to second rank (operating values) using either of these commands:
TRG or T
The device trigger interface command will also shift data from first rank to second rank.
Voltage and current values are compared to soft limits before being loaded into either first or second rank, and new soft limits are compared to values already programmed in both first and second rank. Any discrepancy between first and second rank voltage or current settings and voltage or current soft limits will result in an error condition.
The power supply has been designed to allow it to enforce sequential device command processing, which means that a second instrument on the