Trigger Subsystem

This subsystem controls remote triggering of the power supply.

TRIG[:IMM]

When the trigger subsystem is armed, TRIG generates a trigger signal. The trigger will then:

1.Initiate the output change from the pending level to the immediate level (see CURR[:LEV]:TRIG and VOLT[:LEV]:TRIG).

2.Clear the WTG bit in the Status Operation Condition register.

3.If INIT:CONT has been sent, the trigger subsystem is immediately rearmed for subsequent triggers. This means that as soon as it is cleared, the WTG bit is again set to 1.

Command Syntax

TRIGer[:IMMediate]

Examples

TRIG TRIGGER:IMMEDIATE

TRIG:SOUR

Sets the trigger source to bus commands. Since the power supply has no other trigger source than the GP-IB bus, this command need not be used. It is included in the command set to provide programming compatibility with other instruments (such as the Agilent Electronic Load family) that may have more than one trigger source.

Command Syntax

TRIGer:SOURce BUS

Parameters

<CRD> BUS

*RST Value

BUS

Examples

TRIG: SOUR BUS TRIGGER: SOURCE BUS

Query Syntax

TRIG:SOUR?

Returned Parameters

<CRD> BUS

Voltage Subsystem

This subsystem programs the output voltage of the power supply.

VOLT[:LEV]

Sets the immediate voltage level or the pending triggered voltage level of the power supply. The immediate level is the voltage programmed for the output terminals. The pending triggered voltage level is a stored value that will be transferred to the output when a trigger occurs. If no pending triggered level has been programmed, then the immediate command programs both levels to the same value. Once a triggered level has been programmed, it is unaffected by subsequent changes to the immediate level.

Command Syntax

[SOURce]:VOLTage[:LEVel] [:IMMediate][:AMPLitude] <NRf+>

 

[SOURce]:VOLTage[:lEVel]:TRIGgered[:AMPlitude] <Nrf+>

Parameters

See "VSET" in Table 3-7 of Section III.

Default Sufflx

V

 

Examples

VOLT 200 MV

VOLTAGE:LEVEL 200 MV

 

VOLTAGE:LEVEL:IMMEDIATE:AMPLITUDE 2.5

 

VOLT:TRIG 20

VOLTAGE:LEVEL:TRIGGERED 20

 

VOLTAGE:LEVEL:TRIGGERED:AMPLITUDE 5.5

Standard Commands for Programmable Instruments (SCPI) 103

Page 103