Hardware Errors During Operation. If an error does not occur during selftest but after the power supply has been operating correctly for a time, error -240 is placed in the error queue. Most subsequent commands will not be executed and will also cause one or more - 240 errors to be placed in the queue after each command. If your application requires the controller to be signalled if the power supply fails, program the status registers to allow execution errors or device-dependent errors to generate an SRQ to the controller.

Error Number

-100 -101 -102 -103 -104 -105 -108 -109 -110 -111 -112 -113 -118 -120 -121 -123 -124 -128 -130 -131 -134 -138 -140 -141 -144 -148 -150 -151 -158 -160 -161 -168 -170 -171

Table C-5. Summary of Error Messages

Error String [Description/Explanation/Examples]

Command error [generic] Invalid character

Syntax error [unrecognized command or data type] Invalid separator

Data type error [e.g., "numeric or string expected, got block data"] GET not allowed

Parameter not allowed [too many parameters] Missing parameter [too few parameters] Command header error

Header separator error

Program mnemonic too long [maximum 12 characters] undefined header [operation not allowed]

Query not allowed Numeric data error

Invalid character in number [includes ’’9" in octal data "#q", etc.] Exponent too large [numeric overflow; exponent magnitude > 32 k] Too many digits [number too long; more than 255 digits received] Numeric data not allowed

Suffix error

Invalid suffix [unrecognized units, or units not appropriate] Suffix too long

Suffix not allowed Character data error

Invalid character data [bad character, or unrecognized] Character data too long [maximum length is 12 characters] Character data not allowed

String data error

Invalid string data [e.g., END received before close quote] String data not allowed

Block data error

Invalid block data [e.g., END received before length satisfied] Block data not allowed

Expression error

Invalid expression [e.g., illegal character in expression]