Command

DLY x

DLY xS

DLY xMS

Table 3-7 GP-IB Commands (continued)

 

*Range or **Response

 

Page

To Query

Description

 

 

 

 

*0--31.999 s

Any of these commands is used to program the

57

 

delay time after a new output voltage or current is

 

 

implemented, or RST or OUT ON command is

 

*0--31999 ms

received. During delay power supply disables CV,

 

 

CC, and OR conditions from being labeled as

 

 

faults, and disables foldback protection.

 

 

 

 

DLY?

OUT OFF

OUT 0

OUT ON

OUT 1

OUT?

FOLD OFF,

FOLD 0

FOLD CV

FOLD 1

FOLD CC

FOLD 2

FOLD?

RST

HOLD OFF

HOLD 0

HOLD ON HOLD 1

**DLY xx.xxx

**OUT 0 or **OUT 1

**FOLD 0 or **FOLD 1 or **FOLD 2

Used to read delay time setting.

57

 

 

.Enables or disables power supply output. Power

58

supply remains able to implement commands

 

even while output is disabled

 

Initial Condition: OUT ON

 

 

 

 

Used to read OUTPUT ON/OFF setting

58

 

 

Enables or disables foldback protection which

58

will disable power supply output if power supply

 

switches to whichever mode, CV or CC, is

 

defined as the fold (error) condition. Foldback

 

protection is inhibited during DELAY period.

 

Initial Condition: FOLD OFF.

 

 

 

 

58

Used to read FOLDBACK setting.

 

 

 

Used to reset power supply if output is disabled

59

by overvoltage, remote inhibit, or foldback

 

protection circuits. Power supply resets to present

 

voltage and current settings (values can be

 

changed while unit is disabled).

 

 

 

Determine if certain newly received commands

59

are implemented by power supply upon receipt or

 

are held for later

 

implementation while power supply

 

continues to operate with previously received

 

values. HOLD ON can be used to synchronize

 

power supply changes with actions taken by other

 

devices on the GP-IB. See TRG command. Initial

 

Condition: HOLD OFF

 

 

 

HOLD?

**HOLD 0 or **HOLD 1

Used to read HOLD setting

59

Operating Instructions 51

Page 51
Image 51
Agilent Technologies 603xA, 6035A, 6028A, 6033A, 6023A GP-IB Commands Range or **Response To Query Description, Hold?