Table 1-2. Supplemental Characteristics (continued)
DC Floating Voltage: Either output terminal may be floated up to the following voltage (including the output voltage) from earth ground:
±240 Vdc on Models 6031A, 6032A, 6033A, and 6038A
±550 Vdc on Models 6030A and 6035A
Exceeding these voltage can result in damage to the equipment.
Remote Sensing: The power supply maintains specifications at the load with up to 0.5 volt drop per load lead. Operation with up to 2 volts per load lead is possible with some degradation of the load effect specification.
Absolute maximum ratings
∙between FLT Terminals 1 and 2 is 17.5 Volts dc. (external resistor required to limit current to 1.25 mA max)
∙between INH Terminals 3 and 4 is 5.5 Volts dc.
∙between Terminals 1 or 2 to 4 or chassis to ground is 42 Volts dc.
Fault (FLT) levels (Terminals 1 and 2) (All models)
∙Ioh (Low Level Output Current) is + 1.2 mA maximum
∙Vol (Low Level Output Voltage) is 0.5 Volts Maximum
Inhibit (INH) Levels (Terminals 3 and 4) (All models)
∙Vih (High Level Input Voltage) is 2 Volts
∙Vil (Low Level Input Voltage) 0.5 Volts Maximum
∙tw (Pulse Width) (minimum) = 50 ms (2 ms typ)
∙td (Delay Time) = 2 ms typ
∙Iil (Low Level Input Current) =
Multiple Operations: Up to two similar units may be connected in series or
Temperature Rating (°C):
∙Operating is
∙Storage is - 40 + 75 (all models)
Weight kg. (Ibs)
Model | Agilent 6030A | Agilent 6031A | Agilent6032A | Agilent 6033A | Agilent 6035A | Agilent 6038A |
Net | 16.3 (36) | 17.2 (38) | 16.3 (36) | 9.6 (21) | 16.3 (36) | 9.6 (21) |
Shipping | 21.7 (48) | 22.7 (50) | 21.8 (48) | 11.4 (25) | 21.7 (48) | 11.4 (25) |
Dimensions: See Figure
Certification:
The unit is designed to comply with these requirements:
∙ICE
∙CSA Electrical Bulletin
∙VDE 0871.6.78 Level
∙VDE
∙UL
∙ANSI C39.5 Part 0 Draft