Initiate Subsystem

This subsystem enables the trigger system. When a trigger is enabled, the triggering action will occur upon receipt of a <GET>, *TRG, or TRIGger command. If a trigger circuit is not enabled, all trigger commands are ignored. The actions of the two initiate commands are:

If INIT:CONT is OFF, then INIT[:IMM] arms the trigger system for a single trigger.

If INIT:CONT is ON, then the trigger system is continuously armed and INIT[:IMM] is redundant.

Command Syntax

INITiate[:IMMediate]

 

INITiate:CONTinuous <bool>

Parameters

For INIT[:IMM] (None)

 

For INIT:CONT 0 or OFF 1 or ON

*RST Value

OFF

Examples

INIT INITIATE: IMMEDIATE

 

INIT:CONT 1 INITIATE:CONTINUOUS 1

Query Syntax

For INIT[:IMM] (None)

 

For INIT:CONT INIT:CONT?

Returned Parameters

<NR1> 0 or 1

Measure Subsystem

This query subsystem returns the voltage and current measured at the power supply’s sense

Query Syntax

MEASure:CURRent[:DC?] MEASure:VOLTage[:DC?]

Default Suffix

A for MEAS:CURR

V for MEAS:VOLT

 

Examples

MEAS: CURR? MEAS: VOLT?

 

 

MEASURE: VOLTAGE: DC? MV

 

Returned Parameters

<NR3>

 

Output Subsystem

This subsystem controls the power supply’s voltage and current outputs and the optional Agilent 59510A/59511A Relay Accessory.

OUTP[:STAT]

Enables or disables the power supply’s output. The exact state of a disabled output is model dependent, but it generally is in a condition of minimum output voltage and source current. If the power supply is configured with the Agilent 59510A/59511A Relay Accessory, the command will open the relay contacts when the output is disabled and close them when it is enabled. Use of the second (NORelay) parameter prevents the command from having any effect on the relay; it remains in its existing state when OUTPut is executed. The query form returns the output state, excluding that of the relay (see OUTP:REL?).

Command Syntax

OUTPut[:STATe] <NRf>[,NORelay]

Parameters

{O/OFF}[,NORelay] {I/ON}[,NORelay]

*RST Value

0

Examples

OUTP 1 OUTPUT: STATE ON, NORELAY

Query Syntax

OUTPut[:STATe]?

Returned Parameters

<NRl> 0 or 1

Standard Commands for Programmable Instruments (SCPI) 97

Page 97
Image 97