Section ll Manual Changes

For Agilent Models 6030A, 6031A, 6032A and 6035A: on page 2-7 (AC line Impedance Check, step b), where the maximum output voltages are tabulated, change the voltages as shown below:

6030A change 65 V to 50 V

6031A change 8 V to 6 V

6032A change 22 V to 13.5 V

6035A change 220 V to 150 V

Section lll Manual Changes

On page 3-10 (Overvoltage Protection), change 107% to 90%.

On page 3-11 (Monitor Signals), change “0-5 Volts” to ‘’0-4.5 Volts’ ‘ .

In Table 3-6, change the soft voltage limits as follows:

Model

from:

to:

6030A

204.75

174.03

6031A

20.475

17.40

6032A

61.425

52.21

6035A

511.875

435.10

6033A

20.475

17.40

6038A

61.425

52.21

In table 3-7, GP-IB Commands, change the boxes in the second column (*Range **Response) with voltages tabulated in tables a, b, c, or d as shown below.

Change the column for VSETx, VSETxV, VMAXx, and VMAXxV to set a.

Change the column for VSETxmV, and VMAXxmV to set b.

Change the column for ISETxA, ISETxA, IMAXx, and IMAXxA to set c.

Change the column for ISETxmA, and IMAXxmA to set d.

Agilent Model

6030A

6031A

6032A

6033A

6035A

6038A

Set a

0-175.00 V

020.45V

0-51.495 V

0-17.500 V

0-437.50 V

0-51.495 V

Set b

0-175000 mV

0-20425 mV

0-51495 mV

0-17500 mV

0-437500 mV

0-51495 mV

Set c

0-17.403 A

0-122.85 A

0-51.1875 A

0-30.7125 A

0-5.119 A

0-10.2375 A

Set d

0-17403 mA

0-122850 mA

0-51187.5 mA

0-30712.5 mA

0-5119 mA

0-10237.5 mA