Table B-1. ID Query Response

Agilent Model

Agilent 6030A

Agilent 6031A

Agilent 6032A

Agilent 6033A

Response

ID Agilent 6030A or ID Agilent 6030A, Opt 100

ID Agilent 6031A or ID Agilent 6031A, Opt 100

ID Agilent 6032A or ID Agilent 6032A, Opt 100

ID Agilent 6033A or ID Agilent 6033A, Opt 100

If the power supply responds correctly, the microcomputer, associated circuits and software are almost certainly functioning properly.

The most common cause of failure of the power supply to respond to commands sent from the controller is typing errors. To determine if the power supply has detected an error, send string:

ERR?

and address the power supply to talk. The response should be:

ERR x

where x is a decimal digit from 0 to 8. See Table 3-10 for a description of error codes. If the power supply still fails to respond, check that you are using the appropriate format for sending GP-IB commands from your controller. Check that the GP-IB address switches on the rear panel of the power supply are set for the address being used. Substitute another GP-IB instrument for the suspect power supply to determine that a fault exists in the power supply and not in the controller, interface, or cables.

The following steps check the CV and CC loops and the OVP circuit. Remember that measured responses (VOUT,IOUT,OVP) may vary within specified tolerances from the expected response.

a.To check the CV loop, send string:

VSET 20; ISET 1

and wait one second for power supply to settle. Send string:

STS?

and address the power supply to talk. The response should be:

STS 1

indicating that the power supply is in the CV mode. Send string:

VOUT?

and address the power supply to talk. The response should be;

VOUT 20.000

b.To check the OVP circuit, send string:

VSET 10

To run the following program (Agilent-85 format is used here.

80 Blank Front Panel Option 001

Page 80
Image 80
Agilent Technologies 6035A, 6028A, 6033A, 603xA, 6023A, 6031A Table B-1. ID Query Response Agilent Model, Vset 20 Iset, Sts