Preparing the Analyzer for Use

 

Power Cable and Grounding Requirements

 

On the GPIB connector, pin 12 and pins 18 through 24 are tied to chassis

 

ground and the GPIB cable shield. The instrument frame, chassis, covers,

 

and all exposed metal surfaces including the connectors’ outer shell are

 

connected to chassis ground. However, if channel 2 in the IF section is

 

not installed, the channel 2 BNC connector’s outer shell is not connected

 

to chassis ground.

 

DO NOT interrupt the protective earth ground or ‘’float’’ the analyzer.

Warning

 

This action could expose the operator to potentially hazardous

 

voltages.

 

The analyzer is equipped with two three-conductor power cords which

 

 

ground the analyzer when plugged into appropriate receptacles. The

 

type of power cable plug shipped with each analyzer depends on the

 

country of destination. The following figure shows available power cables

 

and plug configurations.