Preparing the Analyzer for Use
This chapter contains instructions for inspecting and installing the analyzer. This chapter also includes instructions for cleaning the screen, transporting and storing the analyzer.
Power Requirements
The analyzer can operate from a single-phase ac power source supplying voltages as shown in the table. With all options installed, the total power consumption of both sections is less than 1025 VA.
| AC Line Voltage |
| |
Range | Frequency |
90-140 Vrms | 47-63 Hz |
198-264 Vrms | 47-63 Hz |
The line-voltage selector switches are set at the factory to match the most commonly used line voltage in the country of destination; the appropriate fuses are also installed. To check or change either the line-voltage selector switch or the fuse, see the appropriate sections later in this chapter.
Only a qualified service person, aware of the hazards involved, should measure the line voltage.
Before applying ac line power to the analyzer, ensure the line-voltage selector switches are set for the proper line voltage and the correct line fuses are installed in the fuse holders.