To connect the optional minimum loss pad

Preparing the Analyzer for Use

4To configure your analyzer for a keyboard other than U.S. English, press [System Utility] [keyboard type]. Then press the appropriate softkey to select the language.

Configuring your analyzer to use a keyboard other than U.S. English only ensures that the analyzer recognizes the proper keys for that particular keyboard. Configuring your analyzer to use another keyboard does not localize the on-screen annotation or the analyzer’s online HELP facility.

To connect the optional minimum loss pad

1

2

Caution

The minimum loss pad (option 1D7) provides a 50 ohm matched impedance to the analyzer and a 75 matched impedance to the device under test.

Connect the minumum loss pad to the RF section’s INPUT or SOURCE connector.

Connect a 75 cable between the minimum loss pad and the device under test. Use either a 75 type-N cable or the supplied 75 type-N(m)-to-BNC(f) adapter and a 75 BNC cable.

Do NOT connect a 50 cable or adapter to the 75 minimum loss pad. The center pin is larger in a 50 type-N connector than in a 75 type-N connector. Connecting a 50 type-N connector to the 75 minimum loss pad will damage the 75 minimum loss pad.