Agilent Technologies 89441A manual 7Set the IF section’s power switch to on

Models: 89441A

1 92
Download 92 pages 54.15 Kb
Page 56
Image 56
7Set the IF section’s power switch to on.

Preparing the Analyzer for Use

6Set the RF section’s rear panel and front panel power switches to on.

Press the ‘’l‘’ symbol end of the rocker-switches located on the lower right of the rear panel and on the lower left of the front panel. The RF section provides standby power for the high precision frequency reference. The rear-panel line switch interrupts all power including standby power when you press the ‘’O‘’

symbol end of the switch. The front-panel power switch interrupts all power except standby power when you press the ‘’OI ‘’ symbol end of the switch.

7Set the IF section’s power switch to on.

Press the ‘’l‘’ symbol end of the rocker-switch located on the lower left of the front panel. The analyzer requires about 30 seconds to complete its power-on routine.

8Test the electrical performance of the analyzer using the operation verification or the performance tests in chapter 2, ‘’Verifying Specifications’’ in the Installation and Verification Guide.

The operation verification tests verify the basic operating integrity of the analyzer; these tests take about 2.5 hours to complete and are a subset of the performance tests. The performance tests verify that the analyzer meets all the performance specifications; these tests take about 5 hours to complete.