Agilent Technologies 89441A manual 5Start an averaged measurement

Models: 89441A

1 92
Download 92 pages 54.15 Kb
Page 35
Image 35
5Start an averaged measurement:

Characterizing a Filter

4Configure the source and measurement for a frequency response measurement:

Press [Source], [source on], Press [source type], [periodic chirp], Press [Return], (bottom softkey) Press [level], .5, [ Vrms].

Press [Res Bw/Window], [rbw mode arb], Press [main window], [uniform].

Press [Range], [channel both], [ch* single range up-down].

Press [Average], [average on] Press [num averages], 50, [enter], Press [average type], [rms (video)].

Press [Auto Scale].

5Start an averaged measurement:

Press [Meas Restart].

The display should appear similar to that shown below. To learn more about the choices you make in this measurement, display online help for the various keys used (see “Using Online Help” if you are not familiar with how to do this).

Note the distinction between selecting the range (the sensitivity of the analyzer’s input circuitry) and selecting the scale (the position of the data on the display).

Frequency response data displays the output of a device-under-test divided by the input

5-3

Page 35
Image 35
Agilent Technologies 89441A manual 5Start an averaged measurement