To change the IF section’s fuse

Preparing the Analyzer for Use

To change the IF section’s fuse

The fuse is installed at the factory to match the most commonly used line voltage in the country of destination.

1Unplug the power cord from the IF section (the section with ‘’Agilent 89410A’’ silk screened on its lower right rear panel).

2Using a small screw driver, press in and turn the fuse holder cap counter-clockwise. Remove when the fuse cap is free from the housing.

3Pull the fuse from the fuse holder cap.

4To reinstall, select the proper fuse and place in the fuse holder cap.

AC Line Voltage

Voltage

 

Fuse

Range

Frequency

Select

Agilent Part

Type

Switch

Number

90-140 Vrms

47-440 Hz

115

2110-0342

8 A 250 V Normal Blow

198-264 Vrms

47-63 Hz

230

2110-0055

4 A 250 V Normal Blow

5Place the fuse holder cap in the housing and turn clockwise while pressing in.