Agilent Technologies 89441A To connect the analyzer to a LAN, internet protocol address Return

Models: 89441A

1 92
Download 92 pages 54.15 Kb
Page 64
Image 64
To connect the analyzer to a LAN

Preparing the Analyzer for Use

To connect the analyzer to a LAN

Analyzers with option UFG, 4 megabyte extended RAM and additional I/O, have a ThinLAN and AUI (attachment unit interface) port for connecting the analyzer to the LAN (local area network).

1Set the power switch to off ( O ).

2Connect the ThinLAN BNC cable to the ThinLAN port or the appropriate media access unit (MAU) to the AUI port.

3Set the power switch to on ( l ).

4Press the following keys:

[Local/Setup]

[LAN port setup]

[port select ThinLAN (BNC)] or [port select AUI (MAU)] [IP address]

internet protocol address [Return]

[LAN power-on active]

See your LAN system administrator for the internet protocol address. Your LAN system administrator can also tell you if you need to set the gateway address or subnet mask.