To set up a frequency response measurement

 

 

Characterizing a Filter

 

 

 

 

 

To set up a frequency response measurement

 

This measurement can only be performed with a 2-channel analyzer—you must

Note:

 

have option AY7.

 

You must use the source output and the channel 1 and channel 2 inputs on the

 

IF section for network measurements.

 

 

 

1Using a BNC “T” adapter or power splitter and BNC cables, connect the analyzer’s SOURCE to the CHANNEL 1 input directly and to the CHANNEL 2 input through a filter as shown in the illustration below.

2Select the IF baseband receiver mode and initialize the analyzer:

Press [Instrument Mode] [receiver] [IF section (0-10 MHz)]. Press [Preset].

3Configure the analyzer to make two-channel frequency response measurements:

Press [Measurement Data], [freq response].