To change the RF section’s fuse

Preparing the Analyzer for Use

To change the RF section’s fuse

The fuse is installed at the factory to match the most commonly used line voltage in the country of destination.

1Unplug the power cord from the RF section (the section with “Agilent 89431A” silk screened on its lower left rear panel).

2Using a small screw driver, pry open the power selector cover.

3Pull the white fuse holder out of the power selector and remove the fuse from the fuse holder.

4Select the proper fuse and place in the fuse holder.

 

AC Line Voltage

Selector

 

 

Fuse

RF Section

Range

Frequency

Agilent Part

Type

Switch

Number

 

Agilent 89431A

90-110 Vrms

47-63 Hz

100

2110-0381

3 A 250 V Slow Blow

Agilent 89431A

103-140 Vrms

47-63 Hz

120

2110-0381

3 A 250 V Slow Blow

Agilent 89431A

198-242 Vrms

47-63 Hz

220

2110-0304 1.5 A 250 V Slow Blow

Agilent 89431A

216-264 Vrms

47-63 Hz

240

2110-0304

1.5 A 250 V Slow Blow

1Align the white arrow on top of the fuse holder with the white arrow on the power selector cover. All three arrows should point in the same direction. Push the fuse holder into the top slot of the power selector.

2Close the power selector by pushing firmly on the power selector cover.

3Check that the correct line voltage appears through the power selector cover.

7-13