Warning
5
6
7
Preparing the Analyzer for Use
To change the RF section’s line-voltage switch
The
Unplug the power cord from the RF section (the section with “Agilent 89431A” silk screened on its lower left rear panel).
Using a small screw driver, pry open the power selector cover.
Remove the cylindrical line voltage selector.
Position the cylindrical line voltage selector so the required voltage will be facing out of the power selector, then reinstall.
| AC Line Voltage |
|
Range | Frequency | Selector Switch |
|
| |
|
|
|
100 | ||
120 | ||
220 | ||
240 |
Only a qualified service person, aware of the hazards involved, should measure the line voltage.
Check to see that the proper fuse is installed. See ‘’To change the RF section’s fuse.’’
Close the power selector by pushing firmly on the power selector cover. Check that the correct line voltage appears through the power selector cover.