Preparing the Analyzer for Use
To connect the sections
Do NOT use the IF section’s EXT REF OUT connector or optional OVEN REF OUT connector as an external reference output.
1Attach the IF section to the RF section.
If the hardware is not installed, follow the instructions supplied with the Rear Panel Lock Foot Kit. If the hardware is already installed, slide the IF section on top of the RF section making sure the front
2Connect the RF section’s SERIAL 2 port to the IF section’s SERIAL 2 port using the supplied serial interface interconnect cable. Make sure the end of the cable with the EMI suppressor is conected to the IF section.
3Connect the RF section’s OVEN REF OUT connector to the EXT REF IN connector using the supplied coax
If the RF section does not have the OVEN REF OUT connector (option AY4, Delete High Precision Frequency Reference), connect a 1 MHz, 2 MHz, 5 MHz, or 10 MHz sine or square wave, with an amplitude greater than 0 dBm to the RF section’s EXT REF IN connector. For best residual
4Connect the RF section’s 10 MHz REF TO IF SECTION connector to the IF section’s EXT REF IN connector using the supplied