Agilent Technologies DP106, DP240, DP212, DP214, DP110 After Restarting 2.5.1. Windows, Windows XP

Models: DC140 DC110 DC211 DC135 DP105 DP106 DP211 DP235 DP210 DP111 DP1400 DP110 DP214 DP212 DP240

1 66
Download 66 pages 37.45 Kb
Page 24
Image 24
2.5. After Restarting

3.Turn on the power of the crate(s), if present, and then the PC and start the operating system.

NOTE: Acqiris Digitizers are equipped with a LED. If this LED is not glowing orange or red when the power is applied there is a severe problem. Either the module is broken or the necessary voltages for its use are not available.

NOTE: For proper system operation when using the IC200, IC414, or other PCI extension interface to connect a CC10X crate to a remote PC, the crate must be powered on before the PC in order for the PC BIOS to recognize the presence of the CompactPCI crate.

4.Devices that were installed using a previous version of Acqiris Software the instruments in these logical positions will still appear as Unknown Devices. This can be changed to the new Acqiris type category with the Grey Diamond icon by Uninstalling the device and then Installing again. Instructions on this procedure can be found in the ReadMe.txt file in the manuals folder of your Acqiris Software installation.

2.5. After Restarting

2.5.1. Windows 2000

Under Windows 2000, you must login with administrator privileges after the first boot following the hardware installation; the Plug&Play system must have the appropriate privileges to be able to complete your hardware installation successfully. After a successful hardware installation, you will be able to use your Acqiris Digitizer(s) with normal privileges.

At the first boot following the hardware installation, Windows will detect the new hardware and will install the devices automatically. The following image will appear.

NOTE: In some systems an application program (such as AcqirisLive) will not yet work correctly at this point. One additional boot cycle may be needed if this is the first time that a hardware board is being installed.

2.5.2. Windows XP

Under Windows XP, you must login with administrator privileges after the first boot following the hardware installation; the Plug&Play system must have the appropriate privileges to be able to complete your hardware installation successfully. After a successful hardware installation, you will be able to use your Acqiris Digitizer(s) with normal privileges.

If you login with administrator privileges after the first boot following the hardware installation, Windows will detect the new hardware and start the “Found New Hardware Wizard” after a few seconds.

User Manual: Family of 8-bit Digitizers

Page 24 of 66

Page 24
Image 24
Agilent Technologies DP106, DP240, DP212, DP214, DP110, DP1400, DP111, DP210, DP235 After Restarting 2.5.1. Windows, Windows XP