Peripheral Devices

N5102A Digital Signal Interface Module

Table 11-8 Clock Source Settings and Connectors

Clock Source

Softkeys

 

N5102A Module Connection

 

 

 

 

 

 

 

Reference

Clock Ratea

Freq Ref

Ext Clock In

Device Interface

 

Frequency

 

 

 

 

 

 

 

 

 

 

External

 

 

 

 

 

 

 

 

 

Device

 

 

 

 

 

 

 

 

 

Internalb

 

 

 

 

 

 

 

 

a.For the Internal selection, this sets the internal clock rate. For the External and Device selections, this tells the interface module the rate of the applied clock signal.

b.There should be no clock signal applied to the Ext Clock In connector.

6.Press the Clock Phase softkey.

From the menu that appears, you can adjust the phase of the clock relative to the data in

90 degree increments. The selections provide a coarse adjustment for positioning the clock on the valid portion of the data. Selecting 180 degrees is the same as selecting a negative clock polarity.

The 90 degree and 270 degree selections are not available when the clock rate is set below

10 MHz or above 200 MHz. If 90 degrees or 270 degrees is selected when the clock rate is set below 10 MHz or above 200 MHz, the phase will change to 0 degrees or 180 degrees, respectively.

NOTE The clock phase and clock skew may need to be adjusted any time the clock rate setting is changed. Refer to “Clock Timing for Phase and Skew Adjustments” on page 215.

7.Enter the required phase adjustment.

8.Press the Return softkey to return to the clock setup menu.

9.Press the Clock Skew softkey.

This provides a fine adjustment for the clock relative to its current phase position. The skew is a phase adjustment using increments of time. This enables greater skew adjustment capability at higher clock rates. For clock rates below 25 MHz, this softkey is inactive.

The skew has discrete values with a range that is dependent on the clock rate. Refer to “Clock Timing for Phase and Skew Adjustments” on page 215 for more information on skew settings.

10.Enter the skew adjustment that best positions the clock with the valid portion of the data.

11.Press the Clock Polarity Neg Pos softkey to Neg.

This shifts the clock signal 180 degrees, so that the data starts during the negative clock transition. This has the same affect as selecting the 180 degree phase adjustment.

12.Make the clock polarity selection that is required for the device being tested.

13.Press the Return hardkey to return to the first- level softkey menu.

The clock source selection is also reflected in the first- level softkey menu graphic. For example, if the device is the new clock source, the graphic will show that the frequency reference is now connected to the DUT and the DUT has an input clock line going to the N5102A module.

Chapter 11

227

Page 241
Image 241
Agilent Technologies E8257D PSG, E8267D PSG manual Clock Source Settings and Connectors

E8267D PSG, E8257D PSG specifications

Agilent Technologies, a recognized leader in electronic measurement and communications solutions, offers a comprehensive range of signal generators, including the E8257D PSG (Pulsed Signal Generator) and E8267D PSG. These instruments are engineered to meet the demanding requirements of wireless communication, aerospace, defense, and various research applications.

The E8257D PSG is known for its versatility and reliability. It operates within a frequency range of 250 kHz to 40 GHz, making it suitable for a wide array of applications, from signal generation to vector modulation. With an output power capability of up to +30 dBm, it delivers high-quality signals with exceptional precision. Its low phase noise performance is especially critical for applications such as radar and communication system testing, where signal integrity is paramount.

One of the standout features of the E8257D is its advanced modulation capabilities, including analog and digital modulation schemes. This flexibility allows engineers to simulate real-world communications environments accurately. The PSG also features a built-in arbitrary waveform generator that enables users to create complex waveforms tailored to specific testing needs, providing a significant advantage in research and development.

On the other hand, the Agilent E8267D PSG is designed to cater to the needs of users requiring a combined signal generation and analysis solution. With the capability to generate signals from 250 kHz to 67 GHz, the E8267D is ideal for millimeter-wave applications, as well as testing next-generation wireless technologies.

This model includes features such as enhanced phase noise performance and faster switching speed, which are crucial for signal integrity in sophisticated networks. The instrument's intuitive user interface and powerful software integration facilitate effortless operation and automation, thereby improving workflow efficiency.

Both the E8257D and E8267D PSG instruments incorporate cutting-edge technologies such as low-noise microwave and RF components, as well as digital signal processing capabilities. They provide users with enhanced accuracy and reliability in their measurements.

In summary, Agilent Technologies' E8257D and E8267D PSG signal generators represent the pinnacle of precision in signal generation technology. With their extensive feature sets, advanced modulation capabilities, and robust performance specifications, these instruments are invaluable tools for engineers and researchers working across various high-tech industries.