Troubleshooting

Error Messages

CAUTION Carefully read the entire message! It may list additional risks with this procedure.

3.Release the Preset key.

4.To continue with the sequence, press Continue (to abort with no lost files, press Abort).

5.When the sequence concludes, do the following:

a.Cycle power.

Cycling power restores all previously installed options. Because calibration files are restored from EEPROM, you should see several error messages.

b.Perform the DCFM/DCΦM calibration.

Refer to the DCFM/DCΦM Cal softkey description in the E8257D/67D PSG Signal Generators Key Reference.

c.Agilent Technologies is interested in the circumstances that made it necessary for you to initiate this procedure. Please contact us at the telephone number listed at http://www.agilent.com/find/assist. We would like to help you eliminate any repeat occurrences.

Error Messages

If an error condition occurs in the signal generator, it is reported to both the front panel display error queue and the SCPI (remote interface) error queue. These two queues are viewed and managed separately; for information on the SCPI error queue, refer to the E58257D/67D PSG Signal Generators Programming Guide.

NOTE

When there is an unviewed message in the front panel error queue, the ERR annunciator

 

 

appears on the signal generator’s display.

 

 

 

 

 

 

 

 

 

 

 

Characteristic

 

 

Front Panel Display Error Queue

 

 

 

 

 

 

Capacity (#errors)

30

 

 

 

 

 

 

 

Overflow Handling

Circular (rotating).

 

Drops oldest error as new error comes in.

 

 

 

 

 

 

 

 

Viewing Entries

Press: Utility

> Error Info > View Next (or Previous) Error Page

 

 

 

 

 

Clearing the Queue

Press: Utility

> Error Info > Clear Error Queue(s)

 

 

 

 

Unresolved Errorsa

Re- reported after queue is cleared.

 

 

 

 

 

 

 

When the queue is empty (every error in the queue has been read, or the queue is cleared), the

 

No Errors

following message appears in the queue:

 

 

 

0

No Error Message(s) in Queue

 

 

 

 

 

 

 

a.Errors that must be resolved. For example, unlock.

Chapter 12

251

Page 265
Image 265
Agilent Technologies E8257D PSG, E8267D PSG manual Error Messages, Characteristic Front Panel Display Error Queue

E8267D PSG, E8257D PSG specifications

Agilent Technologies, a recognized leader in electronic measurement and communications solutions, offers a comprehensive range of signal generators, including the E8257D PSG (Pulsed Signal Generator) and E8267D PSG. These instruments are engineered to meet the demanding requirements of wireless communication, aerospace, defense, and various research applications.

The E8257D PSG is known for its versatility and reliability. It operates within a frequency range of 250 kHz to 40 GHz, making it suitable for a wide array of applications, from signal generation to vector modulation. With an output power capability of up to +30 dBm, it delivers high-quality signals with exceptional precision. Its low phase noise performance is especially critical for applications such as radar and communication system testing, where signal integrity is paramount.

One of the standout features of the E8257D is its advanced modulation capabilities, including analog and digital modulation schemes. This flexibility allows engineers to simulate real-world communications environments accurately. The PSG also features a built-in arbitrary waveform generator that enables users to create complex waveforms tailored to specific testing needs, providing a significant advantage in research and development.

On the other hand, the Agilent E8267D PSG is designed to cater to the needs of users requiring a combined signal generation and analysis solution. With the capability to generate signals from 250 kHz to 67 GHz, the E8267D is ideal for millimeter-wave applications, as well as testing next-generation wireless technologies.

This model includes features such as enhanced phase noise performance and faster switching speed, which are crucial for signal integrity in sophisticated networks. The instrument's intuitive user interface and powerful software integration facilitate effortless operation and automation, thereby improving workflow efficiency.

Both the E8257D and E8267D PSG instruments incorporate cutting-edge technologies such as low-noise microwave and RF components, as well as digital signal processing capabilities. They provide users with enhanced accuracy and reliability in their measurements.

In summary, Agilent Technologies' E8257D and E8267D PSG signal generators represent the pinnacle of precision in signal generation technology. With their extensive feature sets, advanced modulation capabilities, and robust performance specifications, these instruments are invaluable tools for engineers and researchers working across various high-tech industries.