E9300/1/4/A Power Sensor Specifications
Figure 10 Typical SWR 9 kHz to 6 GHz (25°C ± 10°C) E9304A
Maximum Power
+25 dBm (320 mW) average
+33 dBm peak (2 W) <10∝ s
Maximum DC Voltage
The Agilent E9304A sensor is dc coupled. DC coupling of the input allows excellent low frequency coverage. However, the presence of dc voltages mixed with the signal will have an effect on the accuracy of the power measurement (see graph below).
Caution | DC voltages in excess of the maximum value (5 V) can damage the |
| sensing diode. |
|
|
Maximum dc voltage: 5 Vdc (E9304A only)
35