E9300/1B and H Power Sensor Specifications
|
| X2 |
| Normal |
|
|
| |
|
| Mode |
| Mode |
|
|
| |
|
| Maximum Sensor Power | E9300/1B | E9300/1H | ||||
|
|
|
|
| ||||
High Power |
| 40 ms |
| 70 ms |
| +40 dBm | +20 dBm |
|
|
|
|
| |||||
Path |
|
|
|
|
|
| ||
120 ms |
| 210 ms |
|
| ||||
|
|
|
| +32 dBm | +12 dBm |
| ||
|
|
|
|
| ||||
|
|
|
|
|
|
| ||
|
| 210 ms |
| 400 ms |
|
| ||
|
|
|
| +26 dBm | +6 dBm |
| ||
|
|
|
|
| ||||
|
|
|
|
|
|
| ||
Typical |
| 400 ms |
| 1 s |
|
| ||
|
|
| +20 dBm | 0 dBm |
| |||
|
|
|
| |||||
Settling |
| 40 ms |
| 70 ms |
|
|
|
|
Times |
|
|
| +10 dBm |
| |||
|
|
|
|
|
| |||
|
| 70 ms |
| 120 ms |
|
| ||
|
|
|
| 0 dBm |
| |||
|
|
|
|
| ||||
|
|
|
|
|
|
| ||
|
| 400 ms |
| 1 s |
|
| ||
|
|
|
|
| ||||
|
|
|
|
| ||||
Low Power |
|
|
|
|
|
| ||
| 3.4 s |
| 6.5 s |
|
|
|
| |
|
|
|
|
|
| |||
Path |
|
|
|
|
|
| ||
|
| 6.8 s |
| 13 s |
|
|
|
|
|
|
|
|
|
|
|
|
|
Minimum Sensor Power
Sensor
Dynamic
Range
Figure 20 E9300/1B & H Autofilter, default resolution, 10 dB decreasing power step (not across the switching point)
Calibration Factor and Reflection Coefficient
Calibration Factor (CF) and Reflection Coefficient (Rho) data are provided on a data sheet included with the power sensor. This data is unique to each sensor. If you have more than one sensor, match the serial number on the data sheet with the serial number on the power sensor you are using. The CF corrects for the frequency response of the sensor. The Agilent EPM series power meters automatically read the CF data stored in the sensor and use it to make the corrections.
Reflection Coefficient (Rho, or ρ ) relates to the SWR according to the following formula:
1 + ρ SWR = ------------
1 – ρ