E9300/1/4/A Power Sensor Specifications

Switching Point

The Agilent E-series E9300 power sensors have two paths, a low power path covering -60 dBm to -10 dBm, and a high power path covering -10 dBm to

+20 dBm. The power meter automatically selects the proper power level path. To avoid unnecessary switching when the power level is near the -10 dBm point, Switching Point Hysteresis has been added. This hysteresis causes the low power path to remain selected until approximately -9.5 dBm as the power level is increased, above this power the high power path is selected.The high power path remains selected until approximately -10.5 dBm as the signal level decreases, below this power the low power path is selected.

 

Error

 

 

Offset at Switch Point

≤± 0.5% (≤± 0.02 dB) typical

 

 

Switching Point Hysteresis

0.5 dB typical

Zero Set,

Zero Drift and

Measurement Noise

 

Conditions

Zero Set

Zero Drift2

Measurement

 

(RH)1

Noise3

 

 

 

 

 

 

 

 

Lower Range

15% to 75%

500 pW

150 pW

700 pW

(-60 to -

75% to 95%

500 pW

4,000 pW

700 pW

10 dBm)

 

 

 

 

Upper Range

15% to 75%

500 nW

150 nW

500 nW

(-10 to

75% to 95%

500 nW

3,000 nW

500 nW

+20 dBm)

 

 

 

 

 

 

 

 

 

1.RH is the abbreviation for Relative Humidity.

2.Within 1 hour after zero set, at a constant temperature, after a 24 hour warm-up of the power meter with sensor connected.

3.The number of averages at 16 for Normal mode and 32 for x2 mode, at a constant temperature, measured over a one minute interval and two standard deviations.

39