Service

 

Troubleshooting

 

Troubleshooting information is intended to first isolate the power sensor, the

 

cable, or the power meter as the defective component. When the power sensor

 

is isolated, an appropriate Sensor Module must be used for repair.

 

If error message 241 or 310 is indicated on the power meter, suspect a failed

 

power sensor. If no error message is displayed, but a problem occurs when

 

making a measurement, try replacing the cable from the power meter to the

 

power sensor. If the problem still exists, try using a different power sensor to

 

determine if the problem is in the power meter or in the power sensor.

 

 

Caution

Electrostatic discharge will render the power sensor inoperative. Do

 

not, under any circumstances, open the power sensor unless you and

 

the power sensor are in a static free environment.

 

 

Repair of Defective Sensor

There are no serviceable parts inside the Agilent E-series E9300 power sensors. If the sensor is defective, replace the entire “module” with the appropriate “Restored Sensor Module.”

 

Disassembly Procedure

 

Disassemble the power sensor by performing the following steps:

 

 

Caution

Disassemble the power sensor only in a static free workstation.

 

Electrostatic discharge renders the power sensor inoperative.

 

 