Service
| 
 | Troubleshooting | 
| 
 | Troubleshooting information is intended to first isolate the power sensor, the | 
| 
 | cable, or the power meter as the defective component. When the power sensor | 
| 
 | is isolated, an appropriate Sensor Module must be used for repair. | 
| 
 | If error message 241 or 310 is indicated on the power meter, suspect a failed | 
| 
 | power sensor. If no error message is displayed, but a problem occurs when | 
| 
 | making a measurement, try replacing the cable from the power meter to the | 
| 
 | power sensor. If the problem still exists, try using a different power sensor to | 
| 
 | determine if the problem is in the power meter or in the power sensor. | 
| 
 | 
 | 
| Caution | Electrostatic discharge will render the power sensor inoperative. Do | 
| 
 | not, under any circumstances, open the power sensor unless you and | 
| 
 | the power sensor are in a static free environment. | 
| 
 | 
 | 
Repair of Defective Sensor
There are no serviceable parts inside the Agilent 
| 
 | Disassembly Procedure | 
| 
 | Disassemble the power sensor by performing the following steps: | 
| 
 | 
 | 
| Caution | Disassemble the power sensor only in a static free workstation. | 
| 
 | Electrostatic discharge renders the power sensor inoperative. | 
| 
 | 
 | 