E9300/1/4/A Power Sensor Specifications
% Error |
|
|
|
|
|
|
|
|
|
|
1 |
|
|
|
|
|
|
|
|
|
|
0.8 |
|
|
|
|
|
|
|
|
|
|
0.6 |
|
|
|
|
|
|
|
|
|
|
0.4 |
|
|
|
|
|
|
|
|
|
|
0.2 |
|
|
|
|
|
|
|
|
|
|
0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
0 | 5 | 10 | 15 | 20 | ||||||
|
|
|
|
| Power (dBm) |
|
|
|
|
|
Figure 12 Typical Power Linearity at 25°C, after zero and
| calibration, with associated Measurement Uncertainty | |||||
|
|
|
|
|
|
|
|
| 0 to | 10 to | |||
|
| 0 dBm | 10 dBm | 20 dBm | ||
|
|
|
|
|
|
|
| Measurement | ± 0.9% | ± 0.8% | ± 0.65% | ± 0.55% | ± 0.45% |
| Uncertainty |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Note | If the temperature changes after calibration and you choose not to | |||||
|
| |||||
| table) should be added to the Power Linearity specifications shown |
above. The typical maximum Additional Power Linearity error due to temperature change after calibration at 25°C, for small changes in temperature, is ±0.15 %/°C (valid after zeroing the sensor).
For larger changes refer to the following table.