Agilent Technologies manual E9300/1B To 0 dBm To 10 dBm 10 to 20 dBm 20 to 26 dBm

Models: E9300

1 71
Download 71 pages 61.87 Kb
Page 48
Image 48

E9300/1B and H Power Sensor Specifications

% Error

 

 

 

 

 

 

 

 

1

 

 

 

 

 

 

 

 

0.8

 

 

 

 

 

 

 

 

0.6

 

 

 

 

 

 

 

 

0.4

 

 

 

 

 

 

 

 

0.2

 

 

 

 

 

 

 

 

0

 

 

 

 

 

 

 

 

-0.2

 

 

 

 

 

 

 

 

-0.4

 

 

 

 

 

 

 

 

-0.6

 

 

 

 

 

 

 

 

-0.8

 

 

 

 

 

 

 

 

-1

 

 

 

 

 

 

 

 

-10

-5

0

5

10

15

20

25

30

Power Level (dBm)

Figure 17 E9300B Typical Power Linearity at 25°C, after zero and calibration with associated Measurement Uncertainty

E9300/1B

-6 to 0 dBm

0 to 10 dBm

10 to 20 dBm

20 to 26 dBm

 

 

 

 

 

Measurement

± 0.65%

± 0.55%

± 0.45%

± 0.31%

Uncertainty

 

 

 

 

 

 

 

 

 

See Note on page -49.

48

Page 48
Image 48
Agilent Technologies manual E9300/1B To 0 dBm To 10 dBm 10 to 20 dBm 20 to 26 dBm