Getting Started

 

First check the section labelled '63￿5HYLVLRQ￿. Release A.01.11 or later is

 

required. If your power meter has an earlier release, please contact your

 

nearest Service Office (listed on page -70) to arrange an upgrade.

 

Next check the section labelled 0DLQ￿)￿:￿5HY￿. Release A1.04.00 or later is

 

required for single channel meters; release A2.04.00 or later is required for dual

 

channel meters. For E9300 power sensors with suffix ‘B’ or ‘H’, firmware

 

revision A1.06.00 or later is required for single channel meters; revision

 

A2.06.00 or later is required for dual channel meters. If your power meter has

 

an earlier release, please contact your nearest Agilent Service Office (listed on

 

page -70) to arrange an upgrade.

 

 

Note

You can carry out the firmware upgrade yourself if your power meter

 

has the required. Access http://www.agilent.com/find/powermeters

 

and click onthe link:

 

EPM Series E4418B Single-Channel Power Meter” or

 

EPM Series E4419B Dual-Channel Power Meter”. Click the

 

Software, Firmware and Drivers” link and follow the downloading

 

instructions.

 

 

 

Interconnections and Calibration

 

Connect one end of an Agilent 11730 series sensor cable to the Agilent E-series

 

E9300 power sensor and connect the other end of the cable to the power

 

meter’s channel input. Allow a few seconds for the power meter to download

 

the power sensor’s calibration table.

 

 

Caution

The Agilent 9304A Sensor is DC coupled. DC voltages in excess of the

 

maximum value (5 Vdc) can damage the sensing diode.

 

 

 

 

Note

Ensure power sensors and cables are attached and removed in an

 

indoor environment.

 

 

To carry out a zero and calibration cycle as requested by the power meter proceed as follows:

Ensure the Agilent E-series E9300 power sensor is disconnected from any signal source.

When calibrating Agilent E-series E9300B or E9301B sensors, first remove the attenuator.