Agilent Technologies E9300 manual Cal Factor Uncertainty

Models: E9300

1 71
Download 71 pages 61.87 Kb
Page 42
Image 42

E9300/1/4/A Power Sensor Specifications

Cal Factor Uncertainty

(Low Power Path,-60 to -10 dBm)

Frequency

Uncertainty (25°C±±10°C)

Uncertainty (0°C to 55°C)

 

E9300A

E9301A

E9304A

E9300A

E9301A

E9304A

 

 

 

 

 

 

 

9 kHz to 10 MHz

-

-

±1.7%

-

-

±2.0%

10 MHz to 30 MHz

±1.8%

±1.8%

±1.7%

±2.2%

±2.2%

±2.0%

30 MHz to 500 MHz

±1.6%

±1.6%

±1.7%

±2.0%

±2.0%

±2.0%

500 MHz to 1.2GHz

±1.8%

±1.8%

±1.7%

±2.5%

±2.5%

±2.0%

1.2 GHz to 6 GHz

±1.7%

±1.7%

±1.7%

±2.0%

±2.0%

±2.0%

6 GHz to 14 GHz

±1.8%

-

-

±2.0%

-

-

14 GHz to 18 GHz

±2.0%

-

-

±2.2%

-

-

 

 

 

 

 

 

 

Cal Factor Uncertainty

(High Power Path,-10 to +20 dBm)

Frequency

Uncertainty (25°C±±10°C)

Uncertainty (0°C to 55°C)

 

E9300A

E9301A

E9304A

E9300A

E9301A

E9304A

 

 

 

 

 

 

 

9 kHz to 10 MHz

-

-

±2.0%

-

-

±3.4%

10 MHz to 30 MHz

±2.1%

±2.1%

±2.0%

±4.0%

±4.0%

±3.4%

30 MHz to 500 MHz

±1.8%

±1.8%

±2.0%

±3.0%

±3.0%

±3.4%

500 MHz to 1.2GHz

±2.3%

±2.3%

±2.2%

±4.0%

±4.0%

±3.4%

1.2 GHz to 6 GHz

±1.8%

±1.8%

±1.8%

±2.1%

±2.1%

±2.1%

6 GHz to 14 GHz

±1.9%

-

-

±2.3%

-

-

14 GHz to 18 GHz

±2.2%

-

-

±3.3%

-

-

 

 

 

 

 

 

 

42

Page 42
Image 42
Agilent Technologies E9300 manual Cal Factor Uncertainty