E9300/1B and H Power Sensor Specifications

 

Error

 

 

Offset at Switch Point

≤± 0.5% (≤± 0.02 dB) typical

 

 

Switching Point Hysteresis

0.5 dB typical

E9300/1B

Conditions

Zero Set

Zero Drift2

Measurement

(RH)1

Noise3

 

 

 

 

 

 

 

 

 

Lower Range

15% to 75%

500 nW

150 nW

700 nW

(-30 to

75% to 95%

500 nW

4 W

700 nW

+20 dBm)

 

 

 

 

 

 

Upper Range

15% to 75%

500 W

150 W

500 W

(+20 to

75% to 95%

500 W

3 mW

500 W

+44 dBm)

 

 

 

 

 

 

E9300/1H

 

 

 

 

 

 

Lower Range

15% to 75%

5 nW

1.5 nW

7 nW

(-50 to 0 dBm)

75% to 95%

5 nW

40 nW

7 nW

 

 

 

 

 

 

Upper Range

15% to 75%

5

W

1.5 W

5

W

(0 to +30 dBm)

75% to 95%

5

W

30 W

5

W

 

 

 

 

 

 

 

1.RH is the abbreviation for Relative Humidity.

2.Within 1 hour after zero set, at a constant temperature, after a 24 hour warm-up of the power meter with sensor connected.

3.The number of averages at 16 for Normal mode and 32 for x2 mode, at a constant temperature, measured over a one minute interval and two standard deviations.

52

Page 52
Image 52
Agilent Technologies manual E9300/1B Conditions Zero Set Zero Drift2 Measurement