Quick Start: Learning How to Make Measurements

Learning to Make Reflection Measurements

Step 4. Output measurement results

This step in the procedure shows you how to output the measurement results to a printer.

For in-depth information on creating a hardcopy of the measurement results, refer to the "Printing, Plotting, and Saving Measurement Results" chapter in the User's Guide.

1.Connect a printer to the analyzer as described in “To Configure the Analyzer with Printers or Plotters” on page 1-11.

2. Press Display MORE TITLE ERASE TITLE and then create a title for the measurement, as shown in Figure 2-11:

Use an optional keyboard to type the title, or

Use the front panel knob and the softkey menu to select each letter of the title.

3.Press DONE when you finish creating the measurement title. The title appears on the upper-left corner of the analyzer display.

4.Press Local SYSTEM CONTROLLER to set up the analyzer as the controller. If you are using an GPIB printer, ensure that there is not another controller on the bus. (Note that this step is not required when using parallel or serial printers.)

5.Press Copy PRINT MONOCHROME to create a black and white hardcopy.

NOTE

If you encounter a problem when printing a hardcopy, refer to “To Configure

 

the Analyzer with Printers or Plotters” on page 1-11.

 

 

Figure 2-11 Example Measurement Title

Chapter 2

2-17

Page 50
Image 50
Agilent Technologies 8722ET, ES, 8720ET, 8719ET quick start Example Measurement Title