W-CDMA Measurement Application

Amplitude

Description

Specifications

Supplemental Information

 

 

 

Modulation Accuracy (Composite

 

 

EVM)

 

 

BTS Measurements

 

 

25 dBm MLa ≤ −15 dBm

 

RF input power and attenuation are set to

20 to 30 °C

 

meet the Mixer Level range.

Composite EVM

 

 

Range

0 to 25%

 

Floor

1.6%

 

Accuracy b

±1.0%c

 

 

±0.5%d

At EVM measurement in the range of

 

 

12.5% to 22.5%

Peak Code Domain Error

 

 

Accuracy

±1.0 dB

 

I/Q Origin Offset

 

 

DUT Maximum Offset

 

10 dBc (nominal)

Analyzer Noise Floor

 

50 dBc (nominal)

Frequency Error

 

 

Range

 

±3 kHz (nominal)e

Accuracy

±5 Hz + tfaf

 

Time offset

 

 

Absolute frame offset accuracy

±20 ns

 

Relative frame offset accuracy

 

±5.0 ns (nominal)

Relative offset accuracy

±1.25 ns

 

(for STTD diff mode)g

 

 

a.ML (mixer level) is RF input power minus attenuation.

b.For 16 QAM or 64 QAM modulation, the relative code domain error (RCDE) must be better than 16 dB and 22 dB respectively.

c.The accuracy specification applies when the EVM to be measured is well above the measurement floor. When the EVM does not greatly exceed the floor, the errors due to the floor add to the accuracy errors. The errors due to the floor are noise-like and add incoherently with the UUT EVM. The errors depend on the EVM of the UUT and the floor as follows: error = [sqrt(EVMUUT2 + EVMsa2)] EVMUUT, where EVMUUT is the EVM of the UUT in percent, and EVMsa is the EVM floor of the analyzer in percent. For example, if the EVM of the UUT is 7%, and the floor is 2.5%, the error due to the floor is 0.43%.