Case: Radio Std = 3GPP WCDMA

 

 

Dynamic Range, relative

76.5 dB

83.9 dB (typical)

2.515 MHz offseta d

 

 

Sensitivity, absolute

94.7 dBm

100.7 dBm (typical)

2.515 MHz offsetc

 

 

Accuracy

 

 

2.515 MHz offset

 

 

Relatived

±0.12 dB

 

Absolutee

±1.05 dB

±0.34 dB (95th Percentile 2σ)

20 to 30 °C

 

 

 

 

 

a.The dynamic range specification is the ratio of the channel power to the power in the offset specified. The dynamic range depends on the measurement settings, such as peak power or integrated power. Dynamic range specifications are based on default measurement settings, with detector set to average, and depend on the mixer level. Default measurement settings include 30 kHz RBW.

b.This dynamic range specification applies for the optimum mixer level, which is about

18 dBm. Mixer level is defined to be the average input power minus the input attenuation.

c.The sensitivity is specified with 0 dB input attenuation. It represents the noise limitations of the analyzer. It is tested without an input signal. The sensitivity at this offset is specified in the default 30 kHz RBW, at a center frequency of 2 GHz.

d.The relative accuracy is a measure of the ratio of the power at the offset to the main channel power. It applies for spectrum emission levels in the offsets that are well above the dynamic range limitation.

e.The absolute accuracy of SEM measurement is the same as the absolute accuracy of the spec- trum analyzer. See “Absolute Amplitude Accuracy” on page 34 for more information. The numbers shown are for 0 - 3.6 GHz, with attenuation set to 10 dB.