Agilent EXA Signal Analyzer

Amplitude Accuracy and Range

a.Supplemental information: The amplitude detection linearity specification applies at all levels below 10 dBm at the input mixer; however, noise will reduce the accuracy of low level mea- surements. The amplitude error due to noise is determined by the signal-to-noise ratio, S/N. If the S/N is large (20 dB or better), the amplitude error due to noise can be estimated from the equation below, given for the 3-sigma (three standard deviations) level.

3σ = 3(20dB)log 1 + 10–( (S N + 3dB) ⁄ 20dB)

The errors due to S/N ratio can be further reduced by averaging results. For large S/N (20 dB or better), the 3-sigma level can be reduced proportional to the square root of the number of averages taken.

b.The scale fidelity is warranted with ADC dither set to On. Dither increases the noise level by nominally only 0.24 dB for the most sensitive case (preamp Off, best DANL frequencies). With dither Off, scale fidelity for low level signals, around 60 dBm or lower, will nominally degrade by 0.2 dB.

c.Reference level and off-screen performance: The reference level (RL) behavior differs from some earlier analyzers in a way that makes this analyzer more flexible. In other analyzers, the RL controlled how the measurement was performed as well as how it was displayed. Because the logarithmic amplifier in these analyzers had both range and resolution limitations, this behavior was necessary for optimum measurement accuracy. The logarithmic amplifier in this signal analyzer, however, is implemented digitally such that the range and resolution greatly exceed other instrument limitations. Because of this, the analyzer can make measurements largely independent of the setting of the RL without compromising accuracy. Because the RL becomes a display function, not a measurement function, a marker can read out results that are off-screen, either above or below, without any change in accuracy. The only exception to the independence of RL and the way in which the measurement is performed is in the input atten- uator setting: When the input attenuator is set to auto, the rules for the determination of the input attenuation include dependence on the reference level. Because the input attenuation setting controls the tradeoff between large signal behaviors (third-order intermodulation and compression) and small signal effects (noise), the measurement results can change with RL changes when the input attenuation is set to auto.

d.Mixer level = Input Level Input Attenuator

e.The relative fidelity is the error in the measured difference between two signal levels. It is so small in many cases that it cannot be verified without being dominated by measurement uncertainty of the verification. Because of this verification difficulty, this specification gives nominal performance, based on numbers that are as conservatively determined as those used in warranted specifications. We will consider one example of the use of the error equation to compute the nominal performance.

Example: the accuracy of the relative level of a sideband around 60 dBm, with a carrier at 5 dBm, using attenuator = 10 dB, RBW = 3 kHz, evaluated with swept analysis. The high level term is evaluated with P1 = 15 dBm and P2 = 70 dBm at the mixer. This gives a maximum error within ±0.025 dB. The instability term is ±0.018 dB. The slope term evaluates to ±0.050 dB. The prefilter term applies and evaluates to the limit of ±0.005 dB. The sum of all these terms is ±0.098 dB.

f.Errors at high mixer levels will nominally be well within the range of ±0.045 dB × {exp[(P1 Pref)/(8.69 dB)] exp[(P2 Pref)/(8.69 dB)]} (exp is the natural exponent function, ex). In this expression, P1 and P2 are the powers of the two signals, in decibel units, whose relative power is being measured. Pref is 10 dBm (10 dBm is the highest power for which linearity is specified). All these levels are referred to the mixer level.

40

Chapter 1

Page 40
Image 40
Agilent Technologies N9010A specifications 3σ = 320dBlog 〈 1 + 10- S ⁄ N + 3dB ⁄ 20dB〉

N9010A specifications

Agilent Technologies, now known as Keysight Technologies, offers a wide range of sophisticated measurement instruments, one of which is the N9010A, a high-performance signal analyzer. The N9010A is a versatile tool designed to meet the needs of engineers and researchers dealing with complex signal generation and analysis.

The N9010A signal analyzer is built on the X-Series signal analysis platform, making it an outstanding choice for various applications, including wireless communication, radar, and electronic warfare. Its main features include an extensive frequency range from 9 kHz to 6 GHz, which provides users with the capability to analyze a diverse array of signals.

One of the defining characteristics of the N9010A is its high dynamic range and sensitivity. The instrument has been engineered to ensure that weak signals can be accurately detected in the presence of strong interfering signals. This quality is crucial for applications such as spectrum monitoring and interference analysis.

Equipped with advanced digital signal processing techniques, the N9010A allows for real-time analysis of complex signals. The analyzer supports various modulation techniques and can perform measurements on frequency, time, and phase to provide a comprehensive view of signal performance. Users can leverage its abilities to analyze a wide range of formats including LTE, Wi-Fi, and many others.

Another significant feature of the N9010A is its flexibility. The instrument includes various software options and measurement applications that can be configured to meet specific testing requirements. This adaptability is essential for engineers who need to stay current with industry developments and evolving technologies.

User-friendliness is also a hallmark of the N9010A. The device features an intuitive interface that simplifies the navigation process, making it easier for users to set up tests and perform analyses with reduced training time.

Additionally, the N9010A can be integrated into a broader test environment, allowing for seamless operation within automated measurement systems. This capability enhances productivity and efficiency in design validation and compliance testing.

In summary, the Agilent Technologies N9010A signal analyzer combines advanced technology, high performance, and user-friendly features. Its ability to handle a wide range of applications makes it an essential tool in the toolbox of modern engineers and researchers working in signal analysis, making it a valuable investment in the pursuit of technological advancements.