Agilent Technologies N9010A specifications 208

Models: N9010A

1 262
Download 262 pages 12.84 Kb
Page 208
Image 208

Single Acquisition Combined WLAN Measurement Application

N9077A, Combined WLAN 802.11b or 802.11g-DSSS Measurements

Description

Specifications

Supplemental Information

 

 

 

Absolute

 

 

20 – 30°C

±1.59 dB

 

33 MHz offset

 

 

Dynamic Range, relativef

 

68.3 dB (nominal)

Sensitivity, absolute

 

–80.7 dBm (nominal)

Accuracy

 

 

Relative

±0.68 dB

 

Absolute

 

 

20 – 30°C

±1.60 dB

 

 

 

 

a.The dynamic range specification is the ratio of the channel power to the power in the offset specified. The dynamic range depends on the measurement settings, such as peak power or integrated power. Dynamic range specifications are based on default measurement settings, with detector set to average, and depend on the mixer level. Default measurement settings include 100 kHz RBW.

b.This dynamic range specification applies for the optimum mixer level, which is about –11.77 dBm. Mixer level is defined to be the average input power minus the input attenuation.

c.The sensitivity is specified with 0 dB input attenuation. It represents the noise limitations of the ana- lyzer. It is tested without an input signal. The sensitivity at this offset is specified in the default 100 kHz RBW, at a center frequency of 1 GHz.

d.The relative accuracy is a measure of the ratio of the power at the offset to the main channel power. It applies for spectrum emission levels in the offsets that are well above the dynamic range limitation.

e.This dynamic range specification applies for the optimum mixer level, which is about –12.77 dBm. Mixer level is defined to be the average input power minus the input attenuation.

f.This dynamic range specification applies for the optimum mixer level, which is about –12.0 dBm. Mixer level is defined to be the average input power minus the input attenuation.

Description

Specifications

Supplemental Information

 

 

 

CCK 11 Mbps (DSSS)

 

 

MLa = –10 dBm

 

EQ Off

20 to 30 °C

 

Reference Filter: Gaussian

EVM

 

 

Operating range

 

0.1 to 20% (nominal)

Floor

 

1.54% (nominal)

Accuracyb

 

 

from 1% to 2%

 

±0.9% (nominal)

from 2% to 20%

 

±0.40% (nominal)

 

 

 

208

Chapter 18

Page 208
Image 208
Agilent Technologies N9010A specifications 208