TD-SCDMA Measurement Application

Measurements

Description

Specification

Supplemental Information

 

 

 

Test signal with TS0 active and one

 

±6 Hz + tfa (nominal)

HS-PDSCH in TS0

 

 

 

 

 

a.ML (mixer level) is RF input power minus attenuation.

b.The EVM floor is derived for signal power 20 dBm. The signal has only 1 DPCH or HS-PDSCH in TS0.

c.The accuracy specification applies when the EVM to be measured is well above the measurement floor. When the EVM does not greatly exceed the floor, the errors due to the floor add to the accuracy errors.

The errors due to the floor are noise-like and add incoherently with the UUT EVM. The errors depend on the EVM of the UUT and the floor as follows: error = [sqrt(EVMUUT2 + EVMsa2)] EVMUUT,

where EVMUUT is the EVM of the UUT in percent, and EVMsa is the EVM floor of the analyzer in percent. For example, if the EVM of the UUT is 7%, and the floor is 2.5%, the error due to the floor is 0.43%.

d.The accuracy is derived in the EVM range 0 ~ 18%. We choose the maximum EVM variance in the results as the accuracy.

e.This specifies a synchronization range with Midamble.

f.tfa = transmitter frequency x frequency reference accuracy