Option EMC Precompliance Measurements

Amplitude

Amplitude

Description

Specifications

Supplemental Information

 

 

 

EMI Average Detector

 

Used for CISPR-compliant average

 

 

measurements and, with 1 MHz

 

 

RBW, for frequencies above 1 GHz

Default Average Type

 

All filtering is done on the linear

 

 

(voltage) scale even when the display

 

 

scale is log.

Quasi-Peak Detector

 

Used with CISPR-compliant RBWs,

 

 

for frequencies ≤ 1 GHz

Absolute Amplitude Accuracy for

 

Meets CISPR standardsa

reference spectral intensities

 

 

Relative amplitude accuracy versus

 

Meets CISPR standardsa

pulse repetition rate

 

 

Quasi-Peak to average response ratio

 

Meets CISPR standardsa

Dynamic range

 

 

Pulse repetition rates ≥ 20 Hz

 

Meets CISPR standardsa

Pulse repetition rates ≤ 10 Hz

 

Does not meet CISPR standards in

 

 

some cases with DC pulse excitation.

RMS Average Detector

 

Meets CISPR standardsa

a. CISPR 16-1-1 (2007)