n.3GPP requires the use of a root-raised-cosine filter in evaluating the ACLR of a device. The accuracy of the passband shape of the filter is not specified in standards, nor is any method of evaluating that accuracy. This footnote discusses the performance of the filter in this instru- ment. The effect of the RRC filter and the effect of the RBW used in the measurement interact. The analyzer compensates the shape of the RRC filter to accommodate the RBW filter. The effectiveness of this compensation is summarized in three ways:

White noise in Adj Ch: The compensated RRC filter nominally has no errors if the adjacent channel has a spectrum that is flat across its width.

TOIinduced spectrum: If the spectrum is due to thirdorder intermodulation, it has a dis- tinctive shape. The computed errors of the compensated filter are 0.001 dB for the 100 kHz RBW used for UE testing with the IBW method. It is also 0.001 dB for the 390 kHz RBW used with the Fast method, and 0.000 dB for the 27 kHz RBW filter used for BTS testing with the Filtered IBW method. The worst error for RBWs between these extremes is 0.05 dB for a 330 kHz RBW filter.

rms CW error: This error is a measure of the error in measuring a CWlike

spurious component. It is evaluated by computing the root of the mean of the square of the power error across all frequencies within the adjacent channel. The computed rms error of the compensated filter is 0.012 dB for the 100 kHz RBW used for UE testing with the IBW method. It is 0.034 dB for the 390 kHz RBW used with the Fast method and 0.000 dB for the 27 kHz RBW filter used for BTS testing. The worst error for RBWs between 27 kHz and

470 kHz is 0.057 dB for a 430 kHz RBW filter-like spurious component. It is evaluated by computing the root of the mean of the square of the power error across all frequencies within the adjacent channel. The computed rms error of the compensated filter is 0.012 dB for the 100 kHz RBW used for UE testing with the IBW method. It is 0.034 dB for the 390 kHz RBW used with the Fast method and 0.000 dB for the 27 kHz RBW filter used for BTS testing. The worst error for RBWs between 27 kHz and 470 kHz is 0.057 dB for a 430 kHz RBW filter.