Single Acquisition Combined WLAN Measurement Application

N9077A, Combined WLAN 802.11b or 802.11g-DSSS Measurements

Description

Specifications

Supplemental Information

 

 

 

Currier Suppression

 

 

UUT Maximum Suppression

 

–10 dBc (nominal)

Analyzer Noise Floor

 

–46 dBc (nominal)

Frequency

 

 

Range

 

±100 kHz (nominal)

Accuracy

 

±10 Hz+tfac

a.ML (mixer level) is RF input power minus attenuation

b.The accuracy specification applies when the EVM to be measured is well above the measurement floor. When the EVM does not greatly exceed the floor, the errors due to the floor add to the accuracy errors. The errors due to the floor are noise-like and add incoherently with the UUT EVM. The errors

depend on the EVM of the UUT and the floor as follows: error = sqrt(EVMUUT2 + EVMsa2) - EVMUUT, where EVMUUT is the EVM of the UUT in percent, and EVMsa is the EVM floor of the analyzer in percent.

c. tfa = transmitter frequency × frequency reference accuracy.

Chapter 18