Chapter 1 39
Agilent EXA Signal Analyzer
Amplitude Accuracy and Range
Description Specifications Supplemental Information
Display Scale Switching Uncertainty
Switching between Linear and Log 0 dBa

a. Because Log/Lin and Log Scale Switching affect only the display, not the measurement, they

cause no additional error in measurement results from trace data or markers.

Log Scale Switching 0 dBa
Description Specifications Supplemental Information
Display Scale Fidelityabc
Log-Linear Fidelity (relative to the
reference condition of 25 dBm input
through the 10 dB attenuation, or
35 dBm at the input mixer)
Input mixer leveldLinearity
80 dBm ML 10 dBm ±0.15 dB
ML < 80 dBm ±0.25 dB
Relative FidelityeApplies for mixer leveld range from 10
to 80 dBm, mechanical attenuator only,
preamp off, and dither on.
Sum of the following terms:
high level term Up to ±0.045 dBf
instability term Up to ±0.018 dB
slope term From equationg
prefilter term Up to ±0.005 dBh