Chapter 7 HP 53131A/132A Specifications

Measurement Specifications

Measurement Specifications

Frequency, Period

Channel 1 and 2 Range:

0.1 Hz to 225

MHz

4.44 ns to 10 s

Channel 3 Range:

 

 

 

 

Option 030

100

MHz to 3

GHz

0.33 ns to 10 ns

Option 050

200

MHz to 5

GHz

0.2 ns to 5 ns

Option 124

200

MHz to 12.4 GHz

80 ps to 5 ns

(Period 2 or 3 selectable only via the HP-IB interface)

For Automatic or External Arming:

(and signals < 100 Hz using Timed Arming)

LSD Displayed:

tres

×

Frequency

or

----------Gate-----Time------------

 

 

 

Period

RMS Resolution:

tres2 + (2 × Trigger Error2 )⎞

×

Frequency

-----

-----------------Gate--------------Time---------------------------------

or

 

Period

 

 

 

 

 

HP 53131A

HP 53132A

 

 

 

 

tres

 

tres

 

 

 

 

typical

650 ps

 

200 ps

 

 

 

 

see graphs for worst case resolution performance

 

 

For Automatic Arming:

Gate Time

N

 

 

 

 

= ---------------------------

 

 

 

 

 

Frequency

 

 

where N = 1 for Ch1 or Ch2 Frequency < 1 MHz

 

 

4 for Ch1 or Ch2 Frequency > 1 MHz

 

 

128 for Ch3

 

 

 

 

 

 

 

tacc

Frequency

 

 

 

×

or

Systematic Uncertainty: ± Time Base Error

± ----------------------

 

 

 

 

Gate Time

 

 

Period

 

 

 

 

 

 

 

 

HP 53131A

HP 53132A

 

 

 

 

tacc

 

tacc

 

 

 

 

typical

350 ps

 

100 ps

 

 

 

 

worst case

1.25 ns

 

500 ps

 

 

 

Trigger: Default setting is Auto Trigger at 50%

For Time or Digits Arming:

7

LSD Displayed:

2 2 × tres

tjitter

×

Frequency

-------------------------------------------------------------------------------------------

+ ---------------------------

or

 

 

 

 

Gate Time × Number or Samples

Gate Time

 

Period

RMS Resolution:

4 × tres2 + 2 × Trigger Error2

tjitter

Frequency

-------------------------------------------------------------------------------------------

+ ---------------------------

x

or

Gate Time ×

Number or Samples

 

Gate Time

Period

 

HP 53131A

HP 53132A

 

 

 

tres

tjitter

tres

tjitter

 

 

typical 500 ps

50 ps

225 ps

3 ps

 

 

see graphs for worst case resolution performance

 

 

Number of Samples =

 

 

 

 

 

Gate Time x Frequency

(Frequency < 200 kHz)

 

Gate Time x 200,000

(Frequency > 200 kHz)

 

 

tacc

Frequency

 

 

×

or

Systematic Uncertainty: ± Time Base Error ± ----------------------

 

 

 

Gate Time

 

Period

 

 

 

 

 

 

 

HP 53131A

HP 53132A

 

 

 

 

tacc

tacc

 

 

 

 

typical

100 ps

10 ps

 

 

 

 

worst case

300 ps

100 ps

 

 

Trigger: Default setting is Auto Trigger at 50%

The following graphs may also be used to compute errors for Period Measurements. To find the Period error (

P), calculate the frequency of the

input signal ( F =

1

) and find the frequency error ( F) from the chart. Then, calculate the period error as:

P

=

F

× P

---

------

 

P

 

 

 

F

 

Assembly-Level Service Guide

7-5