Chapter 7 HP 53131A/132A Specifications

Measurement Specifications (Continued)

Measurement Specifications (Continued)

Time Interval

Measurement is specified over the full signal ranges7 of Channels 1 and 2.

Results Range:

–1 ns to 105 s

 

 

LSD:

53131A

53132A

-------------------------------------------------------------------------------------------

 

 

500 ps

150 ps

 

 

 

RMS Resolution:

 

tre2 s + Start Trigger Error2 + Stop Trigger Error2

 

tres

53131A

53132A

-------------------------------------------------------------------------------------------

 

750 ps

300 ps

Systematic Uncertainty:

±(Time Base Error × TI) ± Trigger Level Timing Error ± 1.5 ns Differential Channel Error (HP 53131A)

 

 

 

 

± 900 ps Differential Channel Error (HP 53132A)

Time Interval Delay

After a Time Interval Measurement has begun by satisfying the trigger conditions on Channel 1, the instrument will wait for the user-entered delay time to elapse before the end-of-measurement trigger will be accepted on Channel 2. Please refer to Measurement Arming for additional information.

Frequency Ratio: Ch1Ch2

Ch1Ch3

Ch2Ch1

Ch3Ch1

Measurement is specified over the full signal range of each input.

Results Range:

 

10–10to 1011

 

 

 

 

 

 

 

 

 

‘Auto’ Gate Time:

 

100 ms (or sufficient cycles on Channel 2 or 3 to make a valid measurement, whichever is longer)

LSD:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ratio

1

:

---------------------

--------1-----------------------------

 

 

 

Ratio

1

:

-------------- 1--------------------------------------------

 

2

 

Ch2 Freq × Gate Time

 

 

 

 

 

3

 

Ch3 Freq × Gate Time

 

 

 

Ratio

2

:

---------------------

Ch2-------------Freq--------------------------------

 

 

Ratio

3

:

--------------

-------Ch3-------------Freq--------------------------------

 

 

1

 

(Ch1 Freq)2 × Gate Time

 

 

 

 

 

1

 

(Ch1 Freq)2 × Gate Time

 

 

 

RMS Resolution:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ratio

1

:

-2----×---------1+ (Ch1 Freq × Ch2 Trigger Error)2----------------------------------------------------------------------------------------------

Ratio

1

:

-2----×---------1+ (Ch1 Freq × Ch3 Trigger Error)2----------------------------------------------------------------------------------------------

 

2

 

 

 

Ch2 Freq × Gate Time

 

3

 

 

Ch3 Freq × Gate Time

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2 × Ch2 Freq × 1 + (Ch1 Freq × Ch2 Trigger Error)2

 

 

 

 

 

Ratio

2

:

Ratio

3

:

2 × Ch3 Freq × 1 + (Ch1 Freq × Ch3 Trigger Error)2

 

1

 

 

 

(Ch1 Freq)

2

× Gate Time

--------------------------------------------------------------------

2-----------------------------------------------------------------------

 

 

 

 

 

 

 

1

 

 

(Ch1 Freq)

× Gate Time

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

To minimize relative phase measurement error, connect the higher frequency signal to Channel 1 when possible.

7See Specifications for Pulse Width and Rise/Fall Time measurements for additional restrictions on signal timing characteristics.

7

Assembly-Level Service Guide

7-9