Chapter 7 HP 53131A/132A Specifications
Measurement Specifications (Continued)
Measurement Specifications (Continued)
Time Interval
Measurement is specified over the full signal ranges7 of Channels 1 and 2.
Results Range: |
| |
LSD: | 53131A | 53132A |
| 500 ps | 150 ps |
RMS Resolution: | tre2 s + Start Trigger Error2 + Stop Trigger Error2 | |
tres | 53131A | 53132A |
750 ps | 300 ps | |
Systematic Uncertainty: | ±(Time Base Error × TI) ± Trigger Level Timing Error ± 1.5 ns Differential Channel Error (HP 53131A) | |
|
| ± 900 ps Differential Channel Error (HP 53132A) |
Time Interval Delay
After a Time Interval Measurement has begun by satisfying the trigger conditions on Channel 1, the instrument will wait for the
Frequency Ratio: Ch1⁄Ch2 | Ch1⁄Ch3 | Ch2⁄Ch1 | Ch3⁄Ch1 |
Measurement is specified over the full signal range of each input.
Results Range: |
|
|
|
|
|
|
| |||||
‘Auto’ Gate Time: | 100 ms (or sufficient cycles on Channel 2 or 3 to make a valid measurement, whichever is longer) | |||||||||||
LSD: |
|
|
|
|
|
|
|
|
|
|
|
|
Ratio | 1⁄ | : |
|
| Ratio | 1⁄ | : |
|
| |||
| 2 |
| Ch2 Freq × Gate Time |
|
|
| 3 |
| Ch3 Freq × Gate Time |
|
| |
Ratio | 2⁄ | : |
|
| Ratio | 3⁄ | : |
|
| |||
| 1 |
| (Ch1 Freq)2 × Gate Time |
|
|
| 1 |
| (Ch1 Freq)2 × Gate Time |
|
| |
RMS Resolution: |
|
|
|
|
|
|
|
|
| |||
Ratio | 1⁄ | : | Ratio | 1⁄ | : | |||||||
| 2 |
|
| Ch2 Freq × Gate Time |
| 3 |
| Ch3 Freq × Gate Time | ||||
Ratio | 2⁄ | : | Ratio | 3 | : | 2 × Ch3 Freq × 1 + (Ch1 Freq × Ch3 Trigger Error)2 | ||||||
| 1 |
|
| (Ch1 Freq) | 2 | × Gate Time | ⁄ | |||||
|
|
|
|
|
| 1 |
| (Ch1 Freq ) | ||||
|
|
|
|
|
|
|
|
|
|
| ||
To minimize relative phase measurement error, connect the higher frequency signal to Channel 1 when possible.
7See Specifications for Pulse Width and Rise/Fall Time measurements for additional restrictions on signal timing characteristics.
7